HK-1710 Monochromatic Wavelength Dispersive X-ray Fluorescence Spectrometry Tester
Technical specifications:
Instrument:
1. principle:Single wavelength dispersive X-ray fluorescence
2. Monochromatic Wavelength Dispersive X-ray fluorescence spectrometry
3. applicable:sulfur content in crude oil, gasoline, diesel, naphtha and other kinds of petroleum products.
4. detection range:0.15ppm-15%
5. applicable standard:ASTM D7039
X-ray excitation source
1. 50W Cr target X ray light tube
Optical path
1. All focusing hyperboloid bent crystal HT DCCS(Independent intellectual property rights technology)
Detection
1.Thin beryllium window closed proportional counter,thickness of the beryllium window25um
2. Digital multi-channel pulse height analyzer(Independent intellectual property rights technology)
3. sample cup:10ml,sample amount 2ml
4. Sample cup membrane: Prolene® 4.0µm (0.16mil)
5. sample molding device:ESP(Easy Sample Prepare)device
Software
1. high speed embedded processor
2. Touch screen graphics user interface, all English panel and software
3. Function:working curve, automatic calibration, daily calibration, result query, report print, statistic analysis, data output SD card or USB.
Parameters
1. detection range:0.15ppm-15%
2. lowest detectable limit:sulfur LLD=0.15ppm(200s)
3. detection time:30S-300S
4. repeatability:Sn-1<=0.2ppm(2ppm oil),0.5ppm(10ppm oil),2.5ppm(150ppm gasoline)
5. reproducibility:0.5ppm(2ppm),1.00ppm(10ppm),5ppm(150ppm)
Ambient requirements: temperature:10ºC~30ºC; relative humidity:<80%
Dimension and weight: weight:40KG; dimension:610mm(L)X 460mm(W)X 310mm(H)
Power: input220±10 VAC 10A; power1000W
Functions
1. Solid calibration sample automatic calibration function introduction:
for daily use, whether the accuracy of the instrument is within the error range, on the one hand, can be identified through the analysis of the standard sample of known concentration HK-S1710 can also through the analysis of the solid sample of fixed chloride content determine instrument counting rate attenuation, to determine whether can continue to use the standard curve
after use for a long time, if instrument performance attenuation, calibration can be based on the analysis of solid samples to automatic calibration standard curve, one-time all standard curve correction, easy to use, the results are accurate
2. Sample cup forming device ESP(Easy Sample Prepare)Simple sample cup and film forming device, convenient for sample preparation, ensure the consistency of the sample cup and film forming, reduce the human error, can prepare bulk sample under test, to avoid contamination of the sample film, greatly reduce the cost
3. Data analysis softwareDaily analysis of the data stored in the U disk or SD card, computer version of the software for statistical data on storage or classification management, can analyze the type of sample data over a period of time average, concentration range, etc