Our HK-NIM1 is designed by ASTM D2624 Standard Test Methods for Electrical Conductivity of Aviation and Distillate Fuels , whichcovers the determination of the electrical conductivity of aviation and distillate fuels with and without a static dissipator additive. The test methods normally a measurement of the conductivity when the fuel is uncharged, that is, electrically at rest (known as the rest conductivity).
Suitable scope
The instrument is used for determination of mercaptan sulfur in gasolines, kerosines, aviation turbine fuels, and distillate fuels containing from 0.0003 to 0.01 mass % of mercaptan sulfur. Organic sulfur compounds such as sulfides, disulfides, and thiophene, do not interfere. Elemental sulfur in amounts less than 0.0005 mass % does not interfere. Hydrogen sulfide will interfere if not removed.
Standard
GB/T 1792
ASTM D2624
Technical parameters
1, conductivity test
range: 0 ~ 1999 pS/m
resolution: 1pS/m
accuracy: < 2 %±1 pS/m
measurement error: less than request of ASTM D2624
ambient temperature: 0ºC~40ºC
relative humidity: ≤85%RH
power: 9V
dimension: 80 × 140 × 35(mm)
calibration value: certificate value± 2 pS/m
2, temperature test
temperature range: - 50 ºC~ 50 ºC
resolution: 0.1 ºC
accuracy: < 1 ºC
Features
This tester adopts single chip microcomputer photoelectric isolation technology, stability and repeatability is good, with short polarization, long sampling time, high accuracy, possesses with functions of temperature measurement. Conductivity detection is mainly used for light oil products, also can be used in the detection of fuel additive environmental protection
Configuration: temperature measuring probe, conductivity pool, detector.
Conductivity value: measuring range 1-2000 ps/m, resolution: 1 ps/m